52
| Maintenance
2.
Wash the trough in hot, soapy water. Use a small brush to clean in the
corners.
3.
Rinse the trough thoroughly and allow it to dry completely.
4.
Replace the trough in the microplate carrier.
Clean the Priming Plate
Applies only to Cytation 5 models with a dispenser.
Clean the priming plate regularly to prevent bacteria growth and residue buildup.
Wash the plate in hot, soapy water, using a small brush to clean in the corners. Rinse
thoroughly and allow it to dry completely.
Clean the Dispense Tubes and Injectors
Applies only to Cytation 5 models with a dispenser.
The Cytation 5’s dispense tubes and injectors require routine cleaning, at least
quarterly and possibly more frequently depending on the type of fluids dispensed.
Required Materials
•
Protective gloves
•
Safety glasses
•
Mild detergent
•
Clean, lint-free cotton cloths
•
Deionized or distilled water
•
Stylus (stored in a plastic cylinder affixed to the rear of the dispense module or
reader) (PN 2872304)
Remove the Dispense Tubes and Injector Holders