15: Configure the Dispensers in Gen5 |
21
2.
On the Imaging Configuration tab, set the value of all objectives to
none
, and
then click
Send Values
.
3.
Set the correct values for all objectives, and then click
Send Values
.
4.
Click
Auto Calibration
.
If problems persist, contact BioTek TAC.
15: Configure the Dispensers in Gen5
Before you use the dispenser with a BioTek reader, you must set the calibration
values in Gen5.
1.
In Gen5, go to
System > Instrument Configuration
, select your
instrument, and click
View/Modify
.
2.
Click
Setup
, and then select the
Dispenser 1 tab
.
3.
On the keyboard, press CTRL+SHIFT+M to enter maintenance mode for the
Dispenser 1 window.
4.
Enter the syringe calibration values from the label on the rear of the
dispenser box. (See Install the Dispenser.)
5.
Click
Send Volumes
, and then click
Get Volumes
to verify that the
entered values were sent to the instrument.
6.
Select the
Dispenser 2 tab
, and repeat steps 3 through 5 for Dispenser 2.