| Instrument Testing
Parameter
Default Setting
Step Label:
"Reference well A2"
Read Well:
A2
Wavelength
1
Gain:
150
Integration Time:
0:10.00 MM:SS.ss
Delay after plate movement:
350 msec
Dynamic Range:
Standard
Read Height:
1.00 mm
Read Step 2
Detection Method:
Luminescence
Read Type:
Endpoint
Step Label:
"Background"
Read Wells:
F1..G12
Wavelength
1
Gain:
150
Integration Time:
0:10.00 MM:SS.ss
Delay after plate movement:
350 msec
Dynamic Range:
Standard
Read Height:
1.00 mm
Read Step 3
Detection Method:
Luminescence
Read Type:
Endpoint
Step Label:
"Battery Check"
Read Wells:
A7–A8
Wavelength
1
Gain:
80
Integration Time:
0:01.00 MM:SS.ss
Delay after plate movement:
350 msec
Dynamic Range:
Standard
Read Height:
1.00 mm