Inspect/Clean Mirrors |
47
•
The purpose of the cleaning solvent is only to dissolve any adhesive
contamination that is holding debris on the surface. The towel needs to absorb
both the excessive solvent and entrap the debris so that it can be removed from
the surface. Surface coatings on dichroics are typically less hard than the
substrate. It is reasonable to expect that any cleaning will degrade the surface at
an atomic level. Consideration should be given as to whether the contamination
in question is more significant to the application than the damage that may
result from cleaning the surface. In many cases, the AR coatings that are
provided to give maximum light transmission amplify the appearance of
contamination on the surface.
Materials
•
7/64" hex key
•
Linen or cloth gloves
•
Anhydrous reagent-grade ethanol
•
Kimwipes
•
Magnifying glass
•
100% pure cotton balls (for the polarizing filters)
Procedure
1.
Open the access door on the front of the instrument and slide the filter cube
straight out of its compartment.
2.
Set the filter cube on the work surface. Using a 7/64" hex key, remove the
screw and washer located between the emission filter positions.
3.
Carefully lift the filter cube top from the cube.
4.
Wearing linen or cloth gloves, grasp the mirror by its edges and lift it out of
the cube.
The mirrors are seated on a shelf in the bottom of the cube and are not
secured in place.
Cytation 5
Содержание Cytation 5
Страница 13: ...Installation ...
Страница 31: ...14 Install the Imager Module 19 Cytation 5 ...
Страница 37: ...Getting Started ...
Страница 54: ...42 Getting Started BioTek Instruments Inc ...
Страница 55: ...Maintenance ...
Страница 66: ...54 Maintenance BioTek Instruments Inc ...
Страница 67: ...Instrument Testing ...
Страница 113: ...Specifications ...
Страница 122: ...110 Specifications BioTek Instruments Inc ...