![background image](http://html.mh-extra.com/html/biotek/cytation-5/cytation-5_instructions-for-use-manual_2756068050.webp)
38
| Getting Started
Adjustments made using the B&C (brightness and contrast) and Channel
Shift dialogs are for display purposes only; the changes do not affect the
data measurements from the images.
Analyze Captured and Saved Images
1.
With your captured or saved images displayed in the left panel of the
Capture dialog, click
Review/Save
.
2.
If necessary, adjust the brightness and contrast of the image.
3.
Click
Analyze
.
4.
In the Analyze Tool, select which type of analysis you want to perform and
define the parameters for the analysis, then click
Start
. The results are
displayed in the right pane.
Imaging Module: Experiment Mode
Applies only to models with the imaging module.
The following sections
briefly
describe how to use the Gen5 imaging module in
experiment mode. See the Gen5 Help for more complete instructions and
descriptions of these features.
Perform an Image Read
You can create an Image Read step to be used alone or in combination with the
monochromator- or filter-based optics of the Cytation 5.
1.
Create a new protocol and select the Plate Type.
2.
Create a read step, and select
Image
as the Read Method.
3.
(Optional) Enter a step label or unique name for this step. Data sets of the
reading results will use the label in online views, reports, and export files.
4.
Select the objective.
5.
Define whether to read a full plate or specific wells.
6.
Define up to four channels to read:
a.
Defining the color filter.
b.
Select whether exposure is automatic or defined.
c.
Define which wells are used for auto-exposure for each channel, if
applicable.
d.
Define focus and exposure options, if needed.
7.
Define the horizontal and vertical offset from the center of the well.
8.
Select whether to create a single image for each well, an image montage, or
an image Z-stack.
9.
Define advanced options, if needed.
BioTek Instruments, Inc.
Содержание Cytation 5
Страница 13: ...Installation ...
Страница 31: ...14 Install the Imager Module 19 Cytation 5 ...
Страница 37: ...Getting Started ...
Страница 54: ...42 Getting Started BioTek Instruments Inc ...
Страница 55: ...Maintenance ...
Страница 66: ...54 Maintenance BioTek Instruments Inc ...
Страница 67: ...Instrument Testing ...
Страница 113: ...Specifications ...
Страница 122: ...110 Specifications BioTek Instruments Inc ...