80
| Instrument Testing
Parameter
Default Setting
Read Speed:
Normal
Delay after plate movement: 350 msec
Measurements per data point: 50
Lamp Energy: Low
Dynamic Range: Standard
Read Height:
7.00 mm
Cytation 5_FP.prt
This procedure contains one Read step using filters with Fluorescence
Polarization enabled, inside a Plate Mode block.
Parameter
Default Setting
Detection Method:
Fluorescence polarization
Read Type:
Endpoint
Plate Type:
Costar 96 black opaque (#3915)
Shake Step:
Linear for 15 seconds
Delay Step:
5 seconds after shake
Synchronized Mode:
Plate Mode with Timing Control
Read Wells:
A6–H8
Filter Sets:
1 (filter cube)
Filters:
EX 485/20 nm, EM 528/20 nm
Optics Position:
Top 510 nm
Gain:
Auto, Scale to high Wells, A8, 10000
Read Speed:
Normal
Delay after plate movement: 350 msec
Measurements per data point: 60
Lamp Energy: Low
Read Height:
7.0 mm
Cytation 5_TRF.prt
Parameter
Default Setting
Plate Type:
Costar 96-well white opaque
Delay Step:
3 minutes
Read Step 1
Detection Method:
Time-resolved fluorescence
BioTek Instruments, Inc.