Performance Specifications | 11
Average Residual Volume (Evacuation Efficiency)
plate prior to testing.
Vacuum Filtration Evacuation Efficiency
96-Well Filter Plates
Average increased weight of the plate is ≤ 1.2 grams after
dispensing 300 µL of deionized water per well into a Millipore®
MSHVN4450 96-well 0.45µm plates (PN 98258) and vacuum
aspirated for 30 seconds and blotted on a paper towel.
384-Well Filter Plates
Average increased weight of the plate is ≤ 4.0 grams after
dispensing 80 µL of deionized water per well into a Millipore®
MZFCN0W10 384-well 1.2µm plates (PN 98287) and vacuum
aspirated for 10 seconds and blotted on a paper towel.
Dispense Precision
96-Tube Manifold
≤
3.0% CV when dispensing 300 µL per well of deionized water with
0.1% Tween 20, with FD&C #1 blue dye at rate 6 into a Costar 96-well
flat-bottomed plate. The absorbance of the solution is read at 630 nm
and 450 nm reference.
192-Tube
Manifold
≤
4.0% CV when dispensing 80 µL per well of deionized water with
0.1% Tween 20, with FD&C #1 blue dye at rate 7 into a Costar 384-
well flat-bottomed plate. The absorbance of the solution is read at 630
nm and 450 nm reference.
Verify™ Clog Detection Technology
Timing
A Verify test shall be completed in less than 5 minutes from initiation
until test results are displayed.
Performance
The Verify level sensor measurement shall have a repeatability standard
deviation of
σ
measurement
<0.14 mm (9.0 µL for 8X8 square well plate),
where the
σ
measurement
applies to a relative volume measurement, i.e.
the delta between two volumes.
BioTek Instruments, Inc.
Содержание 405 LS
Страница 1: ...Operator s Manual Microplate Washer 405 LS ...
Страница 2: ......
Страница 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Страница 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Страница 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Страница 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Страница 238: ......