Qualify the Verify™ Technology Sensor | 169
o
Select Ctrl+A and Ctrl+C
o
Select the
From405
tab and paste (Ctrl+V) all the data into the grid beginning at
worksheet cell
A1
.
If the Sigma Error is
greater than
9 µL, the sensor failed the test.
Troubleshoot as follows:
l
Review the test procedure: Was the Verify Test Plate seated correctly during testing on
the washer and the reader, with well A1 in the correct orientation? Was there a long
delay between steps, allowing fluid to evaporate?
l
Inspect the plate for damage. Replace the plate (according to the prescribed
procedure
)
if the plate is cracked, warped, scratched, or otherwise damaged.
l
Inspect the washer components:
o
Is the Verify sensor's collimator clean and properly installed? (Clean the Verify™
Level Sensor on page 133).
o
Have the plate carrier feet been adjusted? Contact BioTek TAC for recalibration
service.
Take corrective action if possible and rerun the test. If it fails again, contact BioTek
TAC for guidance.
BioTek Instruments, Inc.
Содержание 405 LS
Страница 1: ...Operator s Manual Microplate Washer 405 LS ...
Страница 2: ......
Страница 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Страница 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Страница 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Страница 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Страница 238: ......