42 | Chapter 3: Operation
The LHC keeps track of the last-used COM port for an instrument type. For example,
when an EL406 runs a protocol, the LHC logs the COM port used and the next time an EL406
is used, the LHC applies the same COM port setting. You can disable this feature by defining
your
Ports
preference: select
Tools>Preferences>Ports
.
To correct the COM port for the current protocol, click the
Port
link and use the drop-down
list to select the correct value. The LHC stores the COM port value in the protocol file.
With the 405 LS connected to and communicating with the host computer (i.e.
make sure the instrument is turned on and not busy):
1.
Click the
Open
button, locate the
405 LS
folder and click
Open
.
2. Open the
Maintenance
or other folder and select the desired protocol.
3.
Change the COM port if necessary: click
Port
and enter the correct value
or select from the drop-down list.
4.
Click
Settings
, which opens the Instrument Settings dialog.
5. Under
Get settings from:
click the
instrument
link.
6.
Click
Validate
.
A "Validation successful" message is displayed unless the protocol cannot be
run on your instrument. See
LHC Protocols Explained on page 57
.
7.
Save the protocol.
405™ Microplate Washer LS
Содержание 405 LS
Страница 1: ...Operator s Manual Microplate Washer 405 LS ...
Страница 2: ......
Страница 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Страница 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Страница 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Страница 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Страница 238: ......