100 | Chapter 3: Operation
Tubing
Control
Module
Waste Bottle Cap
3
Short single tube, approx.
12”/30.48 cm
Intermediate
Waste In
Nipple with the wide, split-
tube extension in the bottle
n
Two other quite distinct tubing sets are used in subsequent steps.
1. Put the intermediate waste bottle into the bracket on the side of the control
module.
2. Connect the tubing with the Luer fitting to the port labeled
Intermediate Waste
Out
and to the smallest nipple on the waste bottle cap.
3. Connect the longest single tube to the port labeled
Intermediate Vacuum
Supply
and to the nipple without a tube inside the bottle.
4. Connect the shortest tube to the port labeled
Intermediate Waste In
and to the
waste bottle cap nipple with the wide, split-tube extension inside the bottle.
Connect to Waste System
Connecting the Vacuum Filtration module to the washer
405™ Microplate Washer LS