Plate Types and Processing Patterns | 111
Plate Types Table
Plate Type
Columns x
Rows
Plate
Height
Default
Aspirate/
Dispense
Heights
mm
Washer
96 Well
12x8
14.35
29/121
384 Well
24x16
14.22
22/120
384 PCR²
24x16
9.50
2/83
² Low profile 384 PCR plates require manually adjusting the height of the manifold stop screw.
Plate Geometry Diagram
Plate height = physical measurement
Default Dispense Height = (plate height - flange 1.0 mm)
Travel Height = (plate height - flange
Plate Clearance
)
n
If Dispense Height > Travel Height (greater than), the travel height is changed
to match the dispense height.
About Wash Processing Patterns
When using a 96-tube manifold to process a 96-well plate, all wells are processed
simultaneously. To process 384-well plates, a processing pattern is needed, which
also provides the ability to process the plate partially. Some portions of the plate