Chapter 9 - Vector Elite Remote Control
Disabling the Fall Limit
You can also use the remote control to disable the Fall Limit.
1. Select Disabled from the pull-down menu.
2. Press the Save button.
Note: The Save button turns red when the Fall Prevention mode is changed.
Caution: Patient injury could occur when the Fall Limit is disabled, so use extra caution.
Fall Prevention
When a fall is prevented, the trolley locks and goes into Static Unload mode.
1. When the Vector Elite system detects a fall, the Fall Prevented window appear. See Figure 9-11.
2. Press the Winch Up button, to return the patient to standing position. You can also use the
Emergency Lower button to lower the patient to the ground.
3. Press the Close button to close the Fall Prevented window. If the patient is in standing position,
press the DBWS button and (if applicable) the Trolley Tracking button to resume the training
session. There must be at least 10 lbs. (4.5 kgs.) of unloading on the system to activate DBWS
or Trolley Tracking.
Figure 9-11: Fall Prevention Dialog Box
Auto Recovery
To enable or disable Auto Recovery, use the remote control to access the Manual Controls screen. See
Figure 9-3. With auto recovery enabled, if a fall is detected while Auto Recovery is enabled, the system
will maintain rope tension allowing the patient to self-correct and stand up without assistance and resume
the session.