Vector Elite System Technical Specifications
Trolley dimensions
36 inches (91.4 cm) long, 25 inches (63.5 cm)
wide, and 11 inches (27.9 cm) tall
Trolley material
Cover is ABS V0 plastic, base is aluminum grade
60601-T6
Trolley patient weight limit
500 lbs. (227 kgs.)
Support Rope material
8 mm diameter HMPE (high modulus
polyethylene)
Spreader bar material
Anodized aluminum
Environmental conditions for operation
Temperature range: 15C to 35C
Atmospheric pressure range: up to 80kPa
Relative humidity range: 20% to 60%.
Environmental conditions for transport and
storage
Temperature range: -25C to 70C
Atmospheric pressure range: up to 67kPa
Relative humidity range: 5% to 95%
Electrical requirements for single trolley
configured system
110/120VAC, 20A, 50/60Hz or 220/240VAC, 10A
50/60Hz dedicated circuit
Outlet must be capable of delivering 1.8kW
Electrical requirements for multi-trolley
configured system
210/240VAC, 20A, 50/60Hz dedicated circuit
Outlet must be capable of delivering 3.0kW
The system consists of mechanical and electronic components. Inadequate handling of those components
may cause health hazards. Disposal of the system must comply with local regulations.
Содержание VEC-1000
Страница 2: ...Clinician s Guide...
Страница 3: ......
Страница 39: ...30 Clinician s Guide...
Страница 49: ......
Страница 67: ......
Страница 87: ...78 Clinician s Guide...
Страница 103: ......
Страница 107: ...98 Clinician s Guide...