background image

26

Note:

 The autoclavable parts in the 

e

5000 model are the green tip cone (4), 

the black tip ejector assembly and the piston (6).

Note:

 Avoid excess grease. Use only the grease provided with the pipette. 

Note:

 Before reassembling check that no lint or particles are on the surface 

of the piston.

Reassembling:

1 .  Screw the piston (6) clockwise into position.

2. 

e

10, 

e

120, 

e

300, 

e

1000: Place the spring (5) around the piston.

 

e

5000: Attach the green tip cone by screwing it clockwise. Place the 

tip ejector assembly to the tip cone and attach by turning it clockwise. 

Replace the tip cone 

fi

 lter with the tool provided and continue from Step 

5.

3.  Push the tip cone (4) into position and attach it by screwing the tip cone 

holder (3) back. Replace the tip cone 

fi

 lter with the tool provided.

4.  Attach the tip ejector collar (2) by screwing it clockwise.

5.  Replace the battery pack (1). Reset the pipettor and press the start button 

several times to test the movement. Test the tip ejection operation.

Note:

 When reassembling the pipette do not over tighten the parts.

Note:

 It is always necessary to check the performance of the pipette after 

in-house service or maintenance (see paragraph 11).

11.2.1. Multichannel pipettors

Cleaning: 

Use ethanol (70%), isopropanol (60%) or mild detergent and soft lint-free 

cloth to clean the exposed lower parts. Rinse with distilled water if needed 

and let the parts dry.

Autoclaving: 

Dispensing heads of the 

e

10-, 

e

120- and 

e

300 pipettes are autoclavable. 

Dispensing head of the 

e

1200 pipette is not autoclavable. The dispensing 

head is marked either autoclavable- or not autoclavable-symbol (Fig. 10) 

depending on wheather it is autoclavable or not.

Autoclaving instructions:

1.  Remove the tip cone 

fi

 lters if 

fi

 tted

2.   Unscrew the dispensing head of the multichannel 

collar and turning the connecting collar and 

turning the dispensing head counterclockwise 

until it comes off

3.  Sterilize the part in 121°C, 1 bar overpressure for 

20 minutes

Autoclavable

Not autoclavable

(Fig. 10)

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Содержание eLine 730020

Страница 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Страница 2: ...eLINE Instruction Manual Bedienungsanleitung Mode d emploi Manual Usario Instruzioni d impiego san Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisan...

Страница 3: ...Instruction Manual 1 Bedienungsanleitung 35 Mode d emploi 71 Manual Usario 105 Instruzioni d impiego 139 eLINE san Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisan...

Страница 4: ...ipetting P MANUAL 15 8 MULTI MODES 16 8 1 Multi dispensing d 17 8 2 Diluting dd and diluting with mixing dd MIXING 18 8 3 Sequenced dispensing Sd 19 8 4 Multi aspirating SA 20 8 5 Automatic dispensing...

Страница 5: ...protocols typically required by laboratories currently The simple keyboard and easy to read display together with user friendly operation and programming allows these tasks to be performed quickly an...

Страница 6: ...m the type of order However all Biohit eLINE pipette packages contains Pipette Tip Safe Cone Filters and Filter Forceps 10 l Grease in single channel units only Instructions for use Performance certif...

Страница 7: ...attery Cover Charging Contacts Reset Button Finger Hook Tip Cone e5000 Tip Ejector Assembly e5000 4 PIPETTE DESCRIPTION 4 1 Display and controls The control and programming of the Biohit eLINE pipette...

Страница 8: ...piston Press to remove tip Press to reset the movements pipette Mode Ranges and Operating Mode Symbols PROG PROG 1 6 6 storage locations MULTI d multi dispensing dd diluting dd MIXING diluting with mi...

Страница 9: ...piston will stop in the lowest position until the button is released In the operating modes with automatic blow out P dd the piston will automatically return to the home position after the start butto...

Страница 10: ...after 50 250 pipetting cycles and always in case of over aspiration The Biohit Filter Forceps should be used to avoid touching the soiled filters by hand Fig 3 See Chapter 2 1 for ordering informatio...

Страница 11: ...nd China plugs 1 Check that the battery protector is removed and the battery cover is properly closed 2 Place the pipette into the Charging unit Make sure that the metal contacts at the rear of the pi...

Страница 12: ...the pipette 4 The charging time is 12 hours for a completely discharged battery Note It is possible to continue pipetting while the eLINE pipette is charged through the AC adaptor Fig 7 With fully dis...

Страница 13: ...nter key In the shortcut programming the changes are confirmed with the Enter key In the shortcut programming the speed adjustment is not possible The use of the Select key in confirming the setting c...

Страница 14: ...speed excluding speed Prog 1 6 was chosen Start pipetting Store the operating mode with Change selected settings with settings to Prog 1 6 Start pipetting P MANUAL is not available in multichannel mo...

Страница 15: ...with automatic blow out rP Reverse pipetting without blow out P MIXING Pipetting with blow out followed with mixing function P COUNT Pipetting with blow out and displayed cycle counter P MANUAL Manua...

Страница 16: ...p in to the solution then press and hold the START button Mixing is performed automatically as long as the START button is held down 2 Position the tip to dispense Press START button twice Note The mi...

Страница 17: ...e START button to aspirate the next sample 7 3 Pipetting with cycle counter P COUNT The eLINE performs pipetting with blow out and the cycle counter function The cycle counter counts up to 99 pipettin...

Страница 18: ...the manual pipetting mode the aspiration and dispensing are controlled manually The aspiration or dispensing continues as long as the START button is held down within the selected volume range The asp...

Страница 19: ...aton dispensing direction 12 Position the tip to dispense Press the START button 13 When the tip is empty volume reaches 0 l within one second the display will show E Position the tip to empty the tip...

Страница 20: ...and to display the speed out 10 Select the speed out with 1 Min and 9 Max 11 Press or to confirm selections 12 Position the tip to aspirate and press the START button The display shows PRE OUT and ar...

Страница 21: ...play starts to scroll faster 5 Press or to confirm the selection 6 Select the desired sample volume volume 2 with 7 Press to confirm selection and to display the speed in or press and continue from St...

Страница 22: ...e nominal volume of the pipette 1 Press to display MULTI 2 Scroll with until the symbol Sd is displayed 3 Confirm by pressing 4 Select the number of dispensing steps by using 5 Press or to confirm the...

Страница 23: ...e the tip into the liquid again and then release the START button to aspirate the next sample 8 4 Multi aspirating SA The pipette performs consecutive aspiration operations of the programmed volume Th...

Страница 24: ...desired dispensing volume by using the Note When holding down the volume display starts to scroll faster 5 Press or to confirm the selection and to display the aliquots 6 Select the desired number of...

Страница 25: ...ion To continue press the START button immediately after E is displayed and within one second the direction arrow will change Keeping the START button depressed place the tip into the liquid again and...

Страница 26: ...h Biohit eLINE pipettes The operating modes P P MIXING P COUNT dd dd MIXING and SA feature an automatic blow out and others leave an excess liquid in the tip The user should follow the recommendations...

Страница 27: ...xcessive force when mounting the tips as this can hamper the tip ejection Avoid rough handling light pressure is all that is required when using the keyboard or mounting the tips Do not drop the pipet...

Страница 28: ...panol 60 or mild detergent and soft lint free cloth The interior of the tip ejector collar and tip cone can be cleaned with a cotton swab Rinse with distilled water if needed Let the parts dry Grease...

Страница 29: ...the movement Test the tip ejection operation Note When reassembling the pipette do not over tighten the parts Note It is always necessary to check the performance of the pipette after in house service...

Страница 30: ...Stand or Carousel However during long storage periods several months it is recommended to disconnect the Charging Stand and Carousel from the mains outlet It is also advisable to remove the battery pa...

Страница 31: ...with the correction factor Z Z values in the table below Vi mi Z 11 Calculate the mean volume v delivered v Vi 10 12 For conformity evaluation calculate the systematic error es of the measurement in...

Страница 32: ...representative for further information The length of the piston stroke is electronically monitored and the pipette does not normally need recalibration However for special applications the Biohit eLI...

Страница 33: ...ohit eLINE pipettes have a onboard monitoring program to control the performance of each pipetting action If any error message appears to the display it means that the pipette has been unable to perfo...

Страница 34: ...nt of grease on a piston and o ring Grease accordingly Pipette out of established specifications Instrument damaged Return to your Biohit service representative for servicing Pipette blocked aspirated...

Страница 35: ...ipping by the manufacturer The Biohit Quality Assurance Procedure guarantees that the Biohit eLINE electronic pipette you have purchased is ready for use Each Biohit eLINE electronic pipette is CE IVD...

Страница 36: ...0 60 l 0 70 0 30 12 l 2 00 1 50 1 24 730360 8 ch 10 300 l 300 l 0 50 0 20 150 l 0 70 0 30 30 l 2 00 1 00 1 30 730390 8 ch 50 1200 l 1200 l 0 50 0 20 600 l 1 00 0 30 120 l 2 50 1 00 1 24 730420 12 ch 0...

Страница 37: ...device must not be recycled as unsorted municipal waste Instead this device must be collected separately in accordance the local recycling regulations The solid bar used in conjuction with the crossed...

Страница 38: ...oment de la publication et ont t compil es avec le plus grand soin Biohit n est pas responsable des ventuelles erreurs ou omissions et se r serve le droit d apporter des modifications sans pr avis Tou...

Страница 39: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Отзывы: