14
6. Select the speed in with
(1= Min and 9 = Max)
7. Press or
to con
fi
rm selection and to display the speed out
8. Select the speed out with
(1= Min and 9 = Max)
9. Press
or
to con
fi
rm selections
10. Position the tip to aspirate and press the START button
11. Position the tip to dispense. Press the START button
12. Position the tip to discard excess liquid and press the START button
twice
Step 10 Step 11 Step 12
Note:
It is also possible to continue to aspirate and dispense the same volume
without the empty function. To continue, keep the START button depressed
during dispensing and within one second the direction arrow will change.
Continuing to hold the START button down, place the tip into the liquid again
and then release the START button to aspirate the next sample.
7.3. Pipetting with cycle counter (P COUNT)
The
e
LINE performs pipetting with blow-out and the cycle counter function.
The cycle counter counts up to 99 pipettings and then returns back to 0. It
is possible to choose the cycle number to begin with. It is also possible to
change the cycle number or reset the counter during the pipetting by using
the arrow keys.
1. Press to display PIPET
2. Scroll
with until the symbol P and COUNT is displayed.
3. Con
fi
rm by pressing
4. Select the desired pipetting volume by using the
Note:
When holding down
the volume display starts to scroll faster.
5. Press
or
to con
fi
rm selection.
6. Cycle counter will show 1. Use
to select the sample number you want
to begin with (0-99).
7. Press to con
fi
rm the selection and to display the speed in
or press
and continue from Step 12.
▲
▼
▼