30
7. Press to con
fi
rm the recalibration and to display speed in or
press and continue from point 12.
8. Select the speed in with
(1= Min to 9 = Max)
9. Press
or
to con
fi
rm selection and to display the speed out
10. Select the speed out with
(1= Min to 9 = Max)
11. Press
or
to con
fi
rm selections
12. Position the tip to aspirate and press the START button
13. Position the tip to dispense and press the START button
Note:
After recalibration the display will show the operating mode symbol
followed with letters CAL to indicate that the mode and the volume in question
is recalibrated to give out the new volume.
Note:
Changing the mode will reset the recalibration. However, it is possible
to store the recalibrated mode to the storage location in PROG mode as
described in paragraph 8.
13. TROUBLE SHOOTING
The Biohit
e
LINE pipettes have a onboard monitoring program to control the
performance of each pipetting action. If any error message appears to the
display it means that the pipette has been unable to perform the attempted
action properly. In case of an error the display will show Error and PRESS
Ent. message. To clear the message please follow these steps:
1. As this procedure will reset the pipette and the tip ejector please
fi
rst
remove the tip by hand.
2. Recharge the pipettor for 15 minutes.
3. Clear the error message by pressing
. The display will show RESET
and PRESS TIP.
3. Reset the pipette by pressing either of the tip ejector buttons
.
Note:
Repeated occurrence of error messages indicates an internal error
causing failure of the
e
LINE to complete the execution of the pipetting. The
e
LINE therefore needs to be returned to the nearest Biohit Service Centre or
your local Biohit service representative for repair.