21
12. Position the tips to aspirate and press the START button
repeating the
action until the cycle is complete. The total volume is displayed.
13. Position the tips to dispense and press START button twice
.
Step 12 Step 13
8.5. Automatic dispensing (Ad)
The pipette performs repetitive dispensing of a selected volume automatically
at the speci
fi
ed interval (0.1–9.9 seconds). The sum of the dispensing aliquots
and an automatically selected excess volume is aspirated into the tip. The
excess volume is needed to ensure equal operating conditions for each
dispensing step.
1. Press to display MULTI
2. Scroll
with until the symbol Ad is displayed.
3. Con
fi
rm by pressing
4. Select the desired dispensing volume by using the
Note:
When holding down
the volume display starts to scroll faster.
5. Press
or
to con
fi
rm the selection and to display the aliquots
6. Select the desired number of aliquots with
7. Press
or
to con
fi
rm selection
8. Select the desired dispensing pace with
(0.1 - 9.9 seconds)
9. Press to con
fi
rm the selection and to display the speed in or press
and continue from Step 14.
10. Select the speed in with
(1= Min to 9 = Max)
11. Press
or
to con
fi
rm selection and to display the speed out
12. Select the speed out with
(1= Min to 9 = Max)
13. Press
or
to con
fi
rm selections
14. Position the tip to aspirate and press the START button
. The display
shows PRE OUT and arrow down to indicate the reset function.
15. Position the tip to discard priming excess and press the start button
.
▲
▼