33
Volume
Test
Number
of
Cat.No. Channels
Range
Volume
Inaccuracy
Imprecision Dispensings
730080 1-ch 50–1000
μ
l 1000
μ
l 0.40% 0.15%
500
μ
l 0.60% 0.20%
100
μ
l 1.50% 0.50%
50
μ
l 2.00% 1.00% 1–20
730100 1-ch 100–5000
μ
l 5000
μ
l 0.50% 0.15%
2500
μ
l 0.80% 0.20%
500
μ
l 1.00% 0.40% 1–50
730320 8-ch
0.2–10
μ
l 10
μ
l 0.90% 0.50%
5
μ
l 1.50% 0.80%
1
μ
l 4.00% 3.00% 1–50
730340 8-ch
5–120
μ
l 120
μ
l 0.50% 0.20%
60
μ
l 0.70% 0.30%
12
μ
l 2.00% 1.50% 1–24
730360 8-ch 10–300
μ
l 300
μ
l 0.50% 0.20%
150
μ
l 0.70% 0.30%
30
μ
l 2.00% 1.00% 1–30
730390 8-ch 50–1200
μ
l 1200
μ
l 0.50% 0.20%
600
μ
l 1.00% 0.30%
120
μ
l 2.50% 1.00% 1–24
730420 12-ch 0.2–10
μ
l 10
μ
l 0.90% 0.50%
5
μ
l 1.50% 0.80%
1
μ
l 4.00% 3.00% 1–50
730440 12-ch 5–120
μ
l 120
μ
l 0.50% 0.20%
60
μ
l 0.70% 0.30%
12
μ
l 2.00% 1.50% 1–24
730460 12-ch 10–300
μ
l 300
μ
l 0.50% 0.20%
150
μ
l 0.70% 0.30%
30
μ
l 2.00% 1.00% 1–30
730490 12-ch 50–1200
μ
l 1200
μ
l 0.50% 0.20%
600
μ
l 1.00% 0.30%
120
μ
l 2.50% 1.00% 1–24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com