16. To dispense, position the tip, press the START button
once and the
pipettor dispenses the aliquots at the speci
fi
ed pace.
17. Finally position the tip to discard any remaining excess and press the
START button twice
.
Step 14 Step 15 Step 16
Step 17
Note:
It is possible to stop the second counter and interrupt dispensing
at any time by pressing the START button. Releasing the START button
activates the counter again and dispensing continues.
Note:
It is also possible to interrupt the dispensing by pressing the START
button and then ENTER key simultaneously. The display will show E and the
tip can be emptied by pressing the START button.
Note:
Furthermore it is possible to continue to aspirate and dispense the
same volume without the empty function. To continue, press the START
button immediately after E is displayed and within one second the direction
arrow will change. Keeping the START button depressed, place the tip into
the liquid again and then release the START button to aspirate the next
sample.
9. PROG MODE
The Prog mode includes six different storage locations (Prog 1–6). The
favourite operating modes with user-selected settings can be stored to these
locations for future recalls.
9.1. Store operating mode to storage locations:
To store an operating mode to Prog 1–6 locations
1. Complete the programming of the selected operating mode as described
in paragraphs 6 and 7
2. Press
fi
rst
and then simultaneously
to display the PROG 1
3. Choose the desired storage location (Prog 1–6) with
Note:
The display shows simultaneously the mode symbol of the previously
stored mode in the storage location in question.