KBF / KBF P (E5.2) 02/2012
page 48/103
CAUTION
Too high inner chamber temperature.
Damage of the lamps and the charging material.
When operating with illumination, program an additional program section.
Set the safety controller to set-point type “Offset” with set-point 5 °C
9.3 Entry of humidity values
Select the second subroutine “
TP-Program 2
” and confirm by pressing “ENTER”. A program table will
appear, which is initially empty until you enter the humidity values. You can now enter the humidity pro-
gram.
Humidity at the beginning
of the program section
Fan speed in %
(no function)
Program No.
Subprogram TP-Program No. 2
Total number of program
sections
Parameter set (preselected)
Tolerance band limits humidity
(maximum and minimum
humidity)
Operation lines
(no function)
Duration of program
section
Number of start section in case of
repeat cycles
Number of duplicates in case
of repeat cycles
Proceeding further is equivalent to the temperature value entry described in chap. 9.2.
Time course of the subroutines:
When starting the overall program, both subroutines (TP-Program 1 and TP-Program 2) will run synchro-
nously. They should be of the same duration because each of the subroutines becomes inactive after run-
off (i.e., no heating or refrigeration and 50% fan speed after ZP 1 is completed, no humidification after ZP
2 is completed). When the complete program is finished, the controller changes to Idle Mode. Tempera-
ture and humidity proceed towards ambient values.
Entering the set point values for KBF P when operating without illumination:
Increase the setting of the humidity set-point according to the table in chap. 8.2.
Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber
approximates ambient temperature. The fan turns at 50% rate (factory setting).
08:43:55 15.12.09
2
08:43:59 15.12.09
Prog. Select.
Prog 1 PROG 01
Prog 2 PROG 02
Prog 3 PROG 03
Prog 4 PROG 04
Prog 5 PROG 05
Prog 1 PROG 01
TP- Program 2
TP- Program 3
TP- Program 1
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...