KBF / KBF P (E5.2) 02/2012
page 24/103
•
Permissible ambient temperature range during operation
: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C /
77 °F
to which the specified technical data relate. Deviations from the
indicated data are possible for other ambient conditions.
With each degree of ambient temperature >25 °C /
77 °F
, the refrigeration power decreases by
1.5 K.
•
Permissible ambient humidity
: 70 % r.H. max., non-condensing
When operating the chamber at temperature set-points below ambient temperature, high ambient humidi-
ty may lead to condensation on the unit.
•
Installation height: max. 2000 m above sea level.
A water tap (1 bar to 10 bar /
14.5 to 145 PSI)
is necessary for the installation of the humidification system
(chap. 4.3). If no suitable in-house water connection is available, you can manually supply water by filling
the freshwater can (option, chap. 15.9).
When placing several units of the same size side by side, maintain a minimum distance of 250 mm / 9
.84
in
between each unit. Wall distances: rear 100 mm /
3.94 in
, sides 160 mm /
6.29 in
. Spacing above the
unit of at least 100 mm /
3.94 in
, must also be accounted for.
CAUTION
Danger by stacking.
Damage to the units.
∅
Do NOT place climatic chambers on top of each other.
To completely separate the unit from the power supply, you must disconnect the power plug. Install the
unit in a way that the power plug is easily accessible and can be easily pulled in case of danger.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) sev-
eral times a year.
Avoid any conductive dust in the ambiance according to the unit layout complying with pollution degree 2
(IEC 61010-1).
Do not install or operate the climatic chamber KBF / KBF P in potentially explosive areas.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT operate the unit in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.
After turning off the unit, you must close the tap of the water supply. Install the unit in a way that the
freshwater supply is easily accessible.
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...