KBF / KBF P (E5.2) 02/2012
page 47/103
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****.*
Setpoint 1
+100.0
FAN
Operating contacts
Time
00:45:00
Repeat Section
5
Repeat Number
10
Tol.-band min.
-1999.0
Tol.-band max.
+9999.0
Parameter set
1
Setpoint 1
Temperature value at the start of the program section
FAN
Fan speed in %
Operating contacts Operation lines (no function)
Time
Duration of the program section
Repeat Section
No. of start section in case of repeat cycles
Repeat Number
No. of duplicates in case of repeat cycles
Tol.-band min.
Temperature limits (maximum / minimum temperature)
(In case of exceeding: temporary program stop)
Tol.-band max.
Parameter set
Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by pressing “ENTER”.
Then enter the values via the arrow keys, and confirm by pressing “ENTER”.
For a negative set-point entry, enter the numerical value first, and then the minus sign (-).
With set-point type “
Limit
”, the user shall adapt the safety controller (chap. 10.2) or the tem-
perature safety device class 3.3 (option, chap. 10.3) to the highest temperature set-point value
of the program actually used. Check the safety device for each temperature program and
change it if necessary. Set the safety controller set-point or the set-point of the temperature
safety device class 3.3 (option) by approx. 2 °C to 5 °C above the controller temperature set-
point.
Recommended setting: Set-point type “
Offset
” with set-point 2 °C.
Entering the set point values for KBF P when operating without illumination:
Adapt the setting of the temperature set-point according to the table in chap. 8.2.
Performance after completing the program:
The controller changes to Idle Mode. Heating, refrigeration, and humidification are inactive; the chamber
approximates ambient temperature. The fan turns at 50% rate (factory setting).
Note for KBF P:
When the illumination is operating, the inner chamber temperature may rise considera-
bly after the program ends. When operating with illumination, an additional program section at the end of
the program is therefore required. This program section gives the desired final temperature and is repeat-
ed infinitely (enter the number of cycles “Cy” as “-1“), i.e., until cancelling the program manually (press the
EXIT or AUTOMATIC button).
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...