KBF / KBF P (E5.2) 02/2012
page 31/103
4.4 Electrical connection
The climatic chamber comes with a fixed power connection cable that has a length of 1800 mm /
5.9 ft
.
Model
Art. no.
(x = 0 or 1)
Power plug
Voltage
+/-10 %
Current
type
Power
frequency Unit fuse
KBF P 240
KBF P 720
9x20-0244
9x20-0246
Shock-proof
plug
200 V to 240 V
1N~
50Hz
16 Amp
KBF 115
KBF 240
KBF 720
KBF P 240
KBF P 720
9x20-0139
9x20-0150
9x20-0108
9x20-0160
9x20-0143
Shock-proof
plug
200 V to 240 V
1N~
50/60 Hz
16 Amp
KBF 115-UL
KBF 240-UL
KBF 720-UL
KBF P 240-UL
KBF P 720-UL
9x20-0184
9x20-0176
9x20-0168
9x20-0106
9x20-0169
NEMA 6-20P 200 V to 240 V
2~
50/60 Hz
16 Amp
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).
When connecting, please observe the regulations specified by the local electricity supply company and as
well as the VDE directives (for Germany).
•
Pollution degree (acc. to IEC 61010-1): 2
•
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 19.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in case
of danger.
Remark when operating with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
Earth connection essential before connecting supply. Check socket before inserting
plug.
When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5
mAmp is possible. If grounding through the power cable is insufficient or missing, the leakage current may
flow through the user’s body. Correct installation of the professional grade power socket provided by the
user safely avoids this. Before connecting the unit to the socket, please check its grounding contact type
plug for appropriate construction and if it is undamaged.
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...