KBF / KBF P (E5.2) 02/2012
page 53/103
9.7 Repetition of a section or several sections within a program
Here we use the example of a set-point ramp temperature program of chap. 9.5. The shaded sections 02
and 03 shall be repeated e.g. 30 times.
Program
section
Set-point
temp.
Fan
Section
time
Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01
40
100 %
00:30:00
Off
1
0
-1999
+9999
02
60
100 %
01:30:00
Off
1
0
-5
+5
03
90
100 %
01:00:00
Off
1
0
-2
+2
04
90
100 %
03:20:00
Off
1
0
-1999
+9999
05
20
100 %
00:00:01
Off
1
0
-1999
+9999
The following table shows the program that results, whereby the differences to the table above are shad-
ed.
Program
section
Set-point
temp.
Fan
Section
time
Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01
40
100 %
00:30:00
Off
1
0
-1999
+9999
02
60
100 %
01:30:00
Off
1
0
-5
+5
03
90
100 %
01:00:00
Off
2
30
-2
+2
04
90
100 %
03:20:00
Off
1
0
-1999
+9999
05
20
100 %
00:00:01
Off
1
0
-1999
+9999
Sections 02 and 03 will be executed in total 31 times; only then will the program continue.
Entry of the values into the display program table:
To have sections repeated infinitely, enter the number of cycles “Cy” as -1.
9.8 Performance after power failure in Program Mode
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run. The power failure is noted in the event list. No error message is displayed indicat-
ing that a power failure had taken place.
08:43:55 15.12.09
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...