KBF / KBF P (E5.2) 02/2012
page 16/103
1.6 Intended use
Climatic chambers series KBF / KBF P are suitable for exact conditioning of harmless materials. A mixture
of any component of the charging material with air must NOT be explosive. The operating temperature
must lie below the flash point or below the sublimation point of the charging material.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 16) are part of the intended use.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT introduce any substance combustible or explosive at working temperature into
the climatic chamber.
∅
NO explosive dust or air-solvent mixture in the inner chamber.
The charging material shall not contain any corrosive ingredients that may damage the ma-
chine components made of stainless steel, aluminum, and copper. Such ingredients include in
particular acids and halides. Any corrosive damage caused by such ingredients is excluded
from liability by BINDER GmbH.
1.7 Resistance of the humidity sensor against harmful substances
The following list of harmful substances refers only to the humidity sensor and does not include any other
materials incorporated in the unit or prohibited substances in relation to explosion protection.
Some gases - especially clean gases - do not have any influence on the humidity sensor. Others do have
a very small influence, whereas others may influence the sensor to a larger extent.
•
The following gases do not influence the sensor and the humidity measurement: Argon (Ar), carbon
dioxide (CO
2
),helium (He), hydrogen (H
2
), neon (Ne), nitrogen (N
2
), nitrous oxide (N
2
O), oxygen (O
2
)
•
The following gases do not, or to a minor extent influence the sensor and the humidity measurement:
Butane (C
4
H
10
), ethane (C
2
H
6
), methane (CH
4
), natural gas propane (C
3
H
8
)
•
The following gases do not, or to a minor extent influence the sensor and the humidity measurement,
provided that the indicated loads are not exceeded:
Maximum work place
threshold limit value
Tolerated concentration
with permanent load
Substance
Formula
ppm
mg/m
3
ppm
mg/m
3
Ammonia
NH
3
20
14
5500
4000
Acetone
CH
3
COCH
3
500
1200
3300
8000
Benzene
300
1200
150000
Chlorine
Cl
2
0.5
1.5
0.7
2
Acetic acid
CH
3
COOH
10
25
800
2000
Ethyl acetate
CH
3
COOC
2
H
5
400
1400
4000
15000
Ethanol
C
2
H
5
OH
500
960
3500
6000
Ethylene glycol
HOCH
2
CH
2
OH 10
26
1200
3000
Formaldehyde
HCHO
0.3
0.37
2400
3000
Isopropanol
(CH
3
)
2
CHOH
200
500
4800
12000
Methanol
CH
3
OH
200
260
3500
6000
Methyl ethyl ketone
C
2
H
5
COCH
3
200
590
3300
8000
Содержание KBF 115 (E5.2)
Страница 5: ...KBF KBF P E5 2 02 2012 page 5 103 Product registration ...
Страница 98: ...KBF KBF P E5 2 02 2012 page 98 103 19 7 Dimensions KBF 115 ...
Страница 99: ...KBF KBF P E5 2 02 2012 page 99 103 19 8 Dimensions KBF 240 KBF P 240 ...
Страница 100: ...KBF KBF P E5 2 02 2012 page 100 103 19 9 Dimensions KBF 720 KBF P 720 ...