KBF / KBF P (E5.2) 02/2012
page 60/103
10.2.2 Checking and setting safety controller set-point type and safety controller set-
point
Grundeinstellung
User-settings
Instrument data
Contrast
Displ. Power down
Contin. operation
User Level
Safety control. Act + 39,8 °C
Safety control. Set. + 42.0 °C
29
User level code no.
1
(factory setting)
4 x
User Level
Date and time
Summer time
Language
Temperature unit
Buzzer
Safety controller
User-code No.
English
°C
1
Active
Safety Controller
Setpoint type
Setpoint
+ 42.0
Limit
Sollwert Art
Grenzwer
Offset
Setpoint
+0042.0
Sollwert Art
Grenzwer
Offset
Setpoint type
Limit
Offset
In the menu “User Level” select the submenu “Safety controller”.
•
Select the safety controller set-point type “Limit” or “Offset” in the field “Setpoint type”
•
Enter the value for “Limit” or “Offset” in the field “Setpoint”.
For temperature disturbances see alarm indications, chap. 11.