KBF-LQC (E5.2) 12/2010
page 22/106
With temperatures below +3 °C /
37.4°F
, water must be completely removed from the humidifying sys-
tem.
CAUTION
Storage below +3 °C / 37.4°F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the unit.
¾
Contact BINDER Service before any transportation below +3 °C /
37.4°F
.
Permissible ambient humidity:
max. 70 % r.H., non-condensing
CAUTION
Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H.
for a long period.
¾
Dry the appliance completely before shut-down:
•
Set the humidity to 0 % r.H. and turn on humidity switch (4).
•
Set the temperature set point to 60 °C /
140°F
for approx. 2 hours (Manual mode).
•
Only then, shut down the unit at the main power switch (3) and close the tap of the
water supply.
When after storage in a cold location you transfer the unit to its warmer installation site, condensation
may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and
is completely dry.
3.4
Location of installation and ambient conditions
Set up the climatic chamber on a flat, even surface and in a well-ventilated, dry location and align it using
a spirit level. The site of installation must be capable of supporting the unit’s weight (see technical data,
chap. 21.4). The chambers are designed for setting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the unit.
∅
Do NOT set up unit in non-ventilated recesses.
¾
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation
: +18 °C /
64.4°F
to +32 °C /
89.6°F
. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C /
77°F
to which the specified technical data relate. For other ambient condi-
tions, deviations from the indicated data are possible.
With each degree of ambient temperature >25 °C /
77°F
, the refrigeration power decreases by
1.5 K.
•
Permissible ambient humidity
: 70 % r.H. max., non-condensing
When operating the chamber at temperature set-points below ambient temperature, high ambient humid-
ity may lead to condensation on the unit.