KBF-LQC (E5.2) 12/2010
page 53/106
Program entry as set-point step (example of a temperature program)
W/°C
t/min.
01 02 03 04 05 06 07 08
0
20
40
60
80
100
30 120
180
380
Tolerance
minimum
Tolerance
maximum
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan Section
time Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 40
100
%
00:30:00 Off 1 0 -1999 +9999
02 40
100
%
00:00:01 Off 1 0 -1999 +9999
03 60
100
%
01:30:00 Off 1 0 -5
+5
04 60
100
%
00:00:01 Off 1 0 -1999 +9999
05 80
100
%
01:00:00 Off 1 0 -2
+2
06 80
100
%
00:00:01 Off 1 0 -1999 +9999
07 20
100
%
03:20:00 Off 1 0 -1999 +9999
08 20
100
%
00:00:01 Off 1 0 -1999 +9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the program controller MB1:
For rapid transition phases, do NOT program any tolerance limits in order to permit maximum
heating, refrigerating, and humidification speed.
08:43:55 15.12.09