In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the
duration of the power failure. Afterwards, both contacts close automatically.
When using the communication software APT-COM™ 3 DataControlSystem (option, chap.
17.1) via the Ethernet interface of the climatic chamber for data acquisition, the alarm is not
automatically transmitted to the APT-COM™ protocol.
¾
Set the tolerance limits for recording limit value excesses separately in APT-COM™ 3.
17.6 Water protected internal socket (option – not for UL units)
The internal socket is splash proof.
IP system of protection 65 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 50 °C /
122°F
WARNING
Exceeding of the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
∅
Do NOT exceed the temperature set-point of 50 °C /
122°F.
¾
Set the safety controller to 50 °C /
122°F.
¾
With optional temperature safety device, class 3.3, set the mechanical thermostat class
3.1 to 50 °C /
122°F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
and humidity range.
CAUTION
Risk of short circuit.
Damage to the unit.
¾
Use delivered plug only (IP protection type 66). Plug-in the plug and turn it to secure.
¾
If the socket is not used, close the lift-up lid and turn it for securing.