KBF-LQC (E5.2) 12/2010
page 55/106
10.7 Repetition of a section or several sections within a program
The example of a set-point ramp temperature program of chap. 10.5 will serve as an example. The
shaded sections 02 and 03 shall be repeated e.g. 30 times.
Program
section
Set-point
temp.
Fan Section
time Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 40
100
%
00:30:00 Off 1 0 -1999
+9999
02
60
100 %
01:30:00
Off
1
0
-5
+5
03
90
100 %
01:00:00
Off
1
0
-2
+2
04 90
100
%
03:20:00 Off 1 0 -1999
+9999
05 20
100
%
00:00:01 Off 1 0 -1999
+9999
The following table shows the resulting program, whereby the differences to the table above are shaded.
Program
section
Set-point
temp.
Fan Section
time Operation
lines
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 40
100
%
00:30:00 Off 1 0 -1999
+9999
02 60
100
%
01:30:00 Off 1 0 -5 +5
03 90
100
%
01:00:00 Off 2
30 -2 +2
04 90
100
%
03:20:00 Off 1 0 -1999
+9999
05 20
100
%
00:00:01 Off 1 0 -1999
+9999
Sections 02 and 03 will be executed in total 31 times; only then, the program will continue.
Entry of the values into the display program table:
To have sections repeated infinitely, enter the number of cycles “Cy” as -1.
10.8 Performance after power failure in Program Mode
Execution of the program is resumed at the point where the interruption occurred with the latest set-
points reached during program course. The power failure is noted in the event list. No error message is
displayed indicating that a power failure has taken place.