KBF-LQC (E5.2) 12/2010
page 30/106
Remark when operating with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
¾
Earth connection essential before connecting supply. Check socket before inserting
plug.
When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5
mAmp is possible. If grounding through the power cable is insufficient or missing, the leakage current
may flow through the user’s body. Correct installation of the professional grade power socket provided by
the user safely avoids this. Before connecting the unit to the socket, please check its grounding contact
type plug for appropriate construction and if it is undamaged.
4.5
Installation and connection of the light cassettes
The light cassettes can be inserted in
different heights onto the beads at the
lateral walls of the unit. Insert and pull out
the light cassettes only by the handles.
Connect the cables of the light cassettes
to its closest connection socket at the
right side in the back of the chamber.
Put the waterproof plug on the connection
socket. When the plug has engaged, turn
the locking-nut of the plug several times
to the right up to its final stop. The plug is
automatically torn into the socket.
Figure 14: Connecting the light cassettes
Use the delivered covers to protect any unused connection socket.
Do not place any charging material directly onto the light cassettes because those are heated by the fluo-
rescent tubes which would lead to exposing the charging material to undefined temperatures. The tem-
perature directly below or on the light cassettes is not equal to the temperature displayed at the tempera-
ture controller.
Place the charging material on the delivered racks below the light cassettes.
CAUTION
The light cassettes will become hot with temperature set-points >40 °C.
Danger of burning.
∅
Do NOT touch the light cassettes during operation.
¾
Let the light cassettes cool down before changing their position.
When operating the chamber at temperatures > 60 °C /
140°F
, remove the light cassettes.
Otherwise the lifetime of the fluorescent tubes will be considerably reduced.