KBF-LQC (E5.2) 12/2010
page 66/106
(continuation)
X = representing the actual val-
ues of VIS and UVA
LIGHT kLUX W/m
2
X = representing the integrated
values of VIS and UVA
DOSES MLUXh
Wh/m
2
Dose target value LUX reached
LUX DOSE
REACHED
immediately
UVA AND LUX
DOSE REACHED
when also dose target
value UVA is reached
Dose target value UVA reached
UVA DOSE
REACHED
immediately
UVA AND LUX
DOSE REACHED
when also dose target
value LUX is reached
3
Switch “LIGHT INTEGRATOR” in
position “OFF”
LIGHT INTEGR. OFF
4
Maximum temperature with light
sensor of 60 °C reached (only if at
least 1 light sensor is plugged-in)
LIGHT SENSOR 60
°C!
immediately
5
Temperature deviation of more
than +/- 2 °C of the entered set-
point
TEMP RANGE
immediately
TEMP RANGE
after 16 Min.
after 16 min.
Humidity deviation of more than
+/- 5 % r.H. of the entered set-
point
HUMID RANGE
immediately
HUMID RANGE
after 16 min.
after 16 min.
Humidity system turned off with
switch (4)
HUMID OFF
immediately
6
Door open
DOOR OPEN
immediately
DOOR OPEN
after 5 min.
With option locking of the keyboard (BINDER Individual, chap. 17.7):
7
Locked keyboard
KEY LOCK
immediately
Power failure
immediately
The indicated intervals refer to the time after occurrence of the error or notified condition.
Notifying messages from one group (e.g. Group 2) mask the notifying messages of the subsequent
Group (e.g. Group 3). The same is true for the alarm messages of different groups.
Notifying, and alarm messages of Group 3 are displayed in intervals, so messages of subsequent groups
can appear for am moment alternately.
When at least one of the light sensors is plugged-in, the maximum temperature of the chamber is limited
to 60 °C. If this temperature is reached due to an excessively high temperature set-point or to an error,
the message “
LIGHT SENSOR 60 °C!
” is displayed. As soon as the unit has cooled down to a value from
59 °C on, the message disappears.