KBF-LQC (E5.2) 12/2010
page 97/106
21.4 KBF-LQC technical data
Unit size
240
720
Exterior dimensions
Width mm
/
inch
925
/
36.42
1250
/
49.21
Height (incl. castors)
mm /
inch
1460
/
57.48
1925
/
75.79
Depth mm
/
inch
880
/
34.65
887
/
34.92
Depth including door handle, I-triangle, connection
mm /
inch
850
/
33.46
939
/
36.97
Wall clearance rear (minimum) (spacers)
mm /
inch
100
/
3.94
100
/
3.94
Wall clearance side (minimum)
mm /
inch
100
/
3.94
200
/
7.87
Steam space volume
l /
cu.ft.
348 / 12.3
918 /
32.4
Quantity of doors
1
2
Quantity of inner glass doors
1
2
Interior dimensions
Width mm
/
inch
650
/
25.60
973
/
38.31
Height mm
/
inch
785
/
30.91
1250
/
49.21
Depth mm
/
inch
485
/
19.09
576
/
22.68
Interior volume
l /
cu.ft.
247
/
8.7
700
/
24.7
Quantity of racks (standard / max.)
2/7
3/12
Quantity of light cassettes
2
3
Load per rack
kg /
lbs.
30
/
66
45
/
99
Permitted total load
kg /
lbs.
100
/
220
150
/
331
Weight (empty)
kg /
lbs.
215
/
474
376
/
829
Temperature data
(without humidity)
Temperature range without light cassettes 1)
°C /
°F
0 to +70 /
32 to 158
0 to +70 /
32 to 158
Temperature range with light cassettes, with illumination 1)
°C /
°F
+10 to +60 /
50 to 140
+10 to +60 /
50 to 140
Max. heat compensation up to 40 °C /
104°F
, with illumina-
tion
W 400
1000
Climatic data
(with humidity)
Temperature range without light cassettes 1)
°C /
°F
+10 to +70 /
50 to 158
+10 to +70 /
50 to 158
Temperature range with light cassettes, with illumination 1)
°C /
°F
+10 to +60 /
50 to 140
+20 to +60 /
68 to 140
at 25 °C /
77°F
and 60% r.H.
± K
0.2
0.2
Temperature fluctuation
with illumination
at 40 °C /
104°F
and 75% r.H.
± K
0.2
0.2
at 25 °C /
77°F
and 60% r.H.
± K
0.6
1.2
Temperature uniformity
(variation) with illumination at 40 °C /
104°F
and 75% r.H.
± K
0.6
1.2
Humidity range without light cassettes
% r.H.
10 to 80
10 to 80
Humidity range with light cassettes, with illumination
% r.H.
10 to 75
10 to 75
at 25 °C /
77°F
and 60% r.H.
± % r.H.
1.5
≤
2.0
Humidity fluctuation with
illumination
at 40 °C /
104°F
and 75% r.H.
± % r.H.
2
≤
2.0
at 25 °C /
77°F
and 60% r.H.
minutes
3
1
Recovery time after doors
were open for 30 sec. 2)
at 40 °C /
104°F
and 75% r.H.
minutes
4
5
Illumination data per light cassette
Lux 7500 9900
ICH compliant illumination for photo stability testing 3)
UVA W/m
2
1.4
2.1
Содержание APT.line KBF-LQC Series
Страница 102: ...KBF LQC E5 2 12 2010 page 102 106 21 7 Dimensions KBF LQC 240...
Страница 103: ...KBF LQC E5 2 12 2010 page 103 106 21 8 Dimensions KBF LQC 720...
Страница 107: ......