MK (E3) 07/2009
page 35/70
Program table corresponding to the diagram above:
Operation lines
Sk
Program
section
No.
Set-point
temp.
W-1
Fan
FAN
Section
time
Time
3 2 1
Target
section
No
No. of
cycles
Cy
Min.
tolerance
Tmin
Max.
tolerance
Tmax
01 -20
****.
00:30:00
Off
Off
Off
1 0 -1999
+9999
02 -20
****.
00:00:01
Off
Off
Off
1 0 -1999
+9999
03 0
****.
01:30:00
Off
Off
Off
1 0 -5 +5
04 0
****.
00:00:01
Off
Off
Off
1 0
-1999
+9999
05 60
****.
01:00:00
Off
Off
Off
1 0 -2 +2
06 60
****.
00:00:01
Off
Off
Off
1 0 -1999
+9999
07 -40
****.
03:20:00
Off
Off
Off
1 0 -1999
+9999
08 -40
****.
00:00:01
Off
Off
Off
1 0 -1999
+9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the program controller MB1:
PGM
09:17:15 15.12.07
For rapid transition phases, do NOT program any tolerance limits in order to permit maximum
heating speed.
Содержание 9020-0146
Страница 65: ...MK E3 07 2009 page 65 70 18 9 Dimensions MK 115 ...
Страница 71: ......