DD1010 FLYNET
ZERO
Forced zero
(e)
0
Defines the range around zero within which forced zero setting
will be performed on scale unloading or in the presence of a
weight stable signal. "Scale unloading" refers to the moment
when the weight falls below the minimum weighing value. On
indicators subject to metrical verification, this parameter must be
set to 0 (zero).
0.25
0.5
1
2
3
Zero track (e)
0
Defines the range around zero within which zero tracking is
performed to small gradual variations in the weight reading.
Example: snow, dust, water falling on the scale.
For dosing systems, set the value to 0 (zero)
When zero track is disabled, even forced zero is automatically
disabled.
0.25
0.5
1
2
3
Initial reset
Defines the indicator operating mode upon switching on.
ZEROCAM
Upon switching on, the instrument indicates the weight value
present on the scale compared to the zero value programmed at
the time of sampling.
AUTOZ
Upon switching on, the indicator performs the auto-zero. In case
of a legal instrument, the auto-zero occurs in a range between -
5% and +15% of the full scale value around the calibrated zero.
In case of a non-legal instrument this range varies from -50% to
+50%.
Automatic
Zero
ON
The indicator resets the weight if a negative weight reading
remains stable for 5 seconds
OFF