Linearized at one point, Linearized at two
points
In systems which are non-linear, in addition to
zero and full scale calibration, you can also
calibrate intermediate points which are valid
for both the rising and falling sides of the
actual weight/measured weight curve.
For linearized calibration at least 2 sample
weights are required: the first for calibration of
the intermediate point, and a second for
calibration of the full scale value.
For 2-point linearized calibration, at least 3
sample weights are required: the first for
calibration of the first intermediate point, a
second for calibration of the second
intermediate point, and a third for calibration of
the full scale value.
The procedure (indicated by on-screen
instructions) is the same as that used for Zero
Full scale calibration, except that in this case
more points are to be calibrated.
D400
3-31