14. PERSONALIZING METROLOGICAL
PARAMETERS
The terminal setup procedure can be accessed in two different ways.
✔
By pressing
on power up (as explained in the Personalization
chapter of the advanced user manual) it is possible to access
non-metrological parameters. In this way, the metrological parameters
can be displayed but not modified.
✔
With the terminal switched on, after removing the seals, by pressing
the calibration button (see par. 14.1 on page 3-23 ) it is possible to
modify all the parameters, both metrological and non-metrological.
14.1 Calibration button
Key
1. access hole for calibration button
2. seal label area
3. pointed instrument
Figure 14.1 - Accessing the setup procedure by way of the
calibration button (log0123.jpg)
D400
3-23