![Bilanciai D400 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/bilanciai/d400/d400_use-maintenance-and-installation-manual_2750949051.webp)
The program can be copied on an upgrade
board.
Copy programme on
board
Proceed in the following way:
✔
move the jumpers of the upgrade board
from P3-P4 to P1-P2, turning them through
90°;
✔
open the cover of the instrument as shown
in fig. 13.1 on page 3-7 ;
✔
install the modified board as indicated in
fig. 15.3 on page 3-48 ;
✔
connect the power cable and power the
terminal;
✔
press the calibration button;
✔
select the language in the Setup menu;
✔
access
Setup Menu / Maintenance /
Copy programme on board
and
program as instructed by the terminal;
✔
turn off the terminal once the operation has
terminated;
✔
disconnect all cables connected to the
terminal;
✔
remove the board;
✔
move the jumper of the board from P1-P2
to P3-P4, turning them through 90°.