0e, 0.5e, 1e, 1.5e, 2e
forced zero
Defines the range around zero within which
forced zero setting will be performed on scale
unloading or in the presence of a weight stable
signal. "Scale unloading" refers to the moment
when the weight falls below the minimum
weighment value. On terminals subject to
metrical verification, this parameter must be
set to 0 (zero).
Tare device
Activated
It is possible both to set the tare both through
acquisition of the weight on the scale or by
entering a tare value directly from the keypad.
Auto-tare only
It is only possible to set the tare through
acquisition of the weight on the scale; tare
values may not be be entered from the
keypad.
De-activated
Tares may not be set either through
acquisition of the weight on the scale or by
entering a tare value directly from the keypad.
D400
3-28