If you access the setup procedure by pressing the calibration button it is
possible to modify all the terminal parameters.
The correct procedure is as follows:
✔
remove the seals from the terminal (lead or labels) (point 2 fig. 14.1
on page 3-23 );
✔
switch on the terminal;
✔
press the calibration button using a non-metallic pointed instrument
(point 3 fig. 14.1 on page 3-23 ) through the access hole (point 1 of
fig. 14.1 on page 3-23 );
✔
select the menu display language; a few terminal identification
messages will be displayed briefly (program code, version, serial
number) immediately followed by the Setup menu.
On exiting Setup, the display language selected in the personalization
menu will be restored.
WARNING
Access to setup parameters by way of the calibration button
is restricted to specialised personnel only.
Any tampering by unauthorised personnel shall invalidate the
warranty with immediate effect.
The main metrological parameters are described in the following
paragraphs. For personalization of non-metrological parameters refer to
the Personalization chapter of the advanced user manual.