LB 470 Level
5 Electric Installation
56925BA2 Rev. 03, 11/2017
39
General important points for installation
Connect the earth conductor.
Ground the housing.
Please observe the information signs on the devices.
5.1.1
80B
Circuit Breaker
A circuit breaker according to DIN EN 61010-1
•
must be available,
•
must be easily accessible for the maintenance personnel and
•
is to be included in the company-internal documentation.
The master EVUs / slave modules are not equipped with a separate ON/OFF switch
to connect or disconnect the voltage supply. Make sure that the system can be de-
energised via the external power supply.
The circuit breaker can be installed as an automatic fuse or switch and has to com-
ply with the requirements according to IEC 947-1 and IEC 947-3. If a fuse is applied,
it must not be triggered under a current strength of 4 A per device.
IMPORTANT
The circuit breaker must be located near the device and be properly marked as
belonging to it.
5.1.2
81B
Cables and Lines
Lines are to be connected with special care.
Connection lines and routing must comply with the applicable regulations.
When routing the cables, make sure that the cable insulation cannot be me-
chanically damaged by sharp edges or movable metal parts.
Use the approved BERTHOLD cable or a cable with equivalent specifications
for the connection.
For intrinsically safe systems, the detector must be connected to the equipotential
bonding of the system. The detector is connected via a 2-core (2 x 12) cable with
approx. 5 ... 10 mm diameter. A screened cable can be used in systems with ex-
tremely strong electrical noise. The screen may only be laid out on one side of the
detector. The maximum cable length depends on the cable resistance, which may
not exceed a total (there and back) of 40 ohms. For standard cables from
BERTHOLD TECHNOLOGIES (Id. no. 32024), this results in a cable length of 1000m,
from the evaluation unit to the detector. For intrinsically safe systems, the maxi-
mum permissible inductance and capacitance of the cable should be considered to
the max. 40 ohm.
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...