LB 470 Level
7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
129
Signal Condition: Rapid Switch
Device settings | Setup | Signal Condition | Rapid Switch
IMPORTANT
The use of the function "Rapid Switch" is recommended only for special appli-
cations where the output signal has to adapt rapidly to the new value, e.g. in
case of measurements on small tanks and if sudden level changes occur.
When Rapid Switch (Fig. 108, item 1) is activated, there is a rapid reaction (Fig. 108,
item 2) to a quick change in level. The time constant in this case is set to 1/10. After
that, the time constant is reset to the original value. A change in level is considered
to be a "rapid" change when the count rate is changing more than the entered
sigma value within a short time. The sigma value can be adjusted to the process. A
sigma of 4.0 is factory set at delivery.
1
2
Activate check box Rapid Switch Enabled
Input field Rapid Switch Sigma
Fig. 111 Signal Condition (Rapid Switch)
NOTICE
Rapid Switch and Interference detection must not be activated simultaneously.
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...