![Berthold Duo Xpert Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/berthold/duo-xpert/duo-xpert_operating-manual_2739249106.webp)
7 Main Menu Device Setup
LB 470 Level
106
56925BA2 Rev. 03, 11/2017
Calibration Settings: Table (2 Point Exp. calibration type)
Device settings | Setup | Calibration | Calibration Settings
1
2
3
4
5
6
Table calibration
Button
<Edit>
Button Add measuring point
< + >
Button Completely delete points
<Clear All>
Button Delete point
< - >
Date of the last editing
7
8
9
10
11
Count of the calibration point
Read-In Time [s]
Count rate [cps]
Process value [%]
Determine new calibration point
Fig. 86
Calibration parameters (Table 2 Point Exp. calibration type)
Add new calibration point
1.
Click on the button
< + >
(Fig. 86, item 3).
The window
″
New calibration point
″
(Fig. 86, item 11) opens.
2.
Click on the input field
″
Process value
″
(Fig. 86, item 9) to open the input
field.
3.
Specify the current, actual level in percent and confirm with the Enter key.
4.
Click on the input field
″
Read-In Time
″
(Fig. 86, item 10) to open the input
field.
5.
Specify a measurement time and confirm with the Enter key.
6.
Click on the button
< Start >.
The Window “New calibration Point” (Fig. 86, item 11) opens and the count
rate is determined.
7.
Click on
<OK>
to confirm the new calibration point.
The new calibration point is recorded in the table.
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...