7 Main Menu Device Setup
LB 470 Level
114
56925BA2 Rev. 03, 11/2017
more accurate the result.
4.
Click on the button
< Perform adjust now >
(Fig. 94, item 1).
A new window opens.
5.
Click on the button
< Start >
(Fig. 94, item 5)
.
The measurement starts.
6.
Click
< Ok >
to accept the value.
The level has been adjusted to the process value (see standard display).
IMPORTANT
When executing an Adjust, the measurement parameter set was overwritten.
If the system should be recalibrated, a Recall of the measurement parameter
set to the calibration parameter must be performed. Otherwise the Adjust will
become lost.