LB 470 Level
9 Troubleshooting
56925BA2 Rev. 03, 11/2017
165
M107
Error in the internal real
time clock
M
Malfunction of the real-time clock.
Check Date and Time. If the event oc-
curs frequently, contact Berthold
Service.
M108
CPU temperature sensor
M
The temperature sensor of the device is
defective. Contact Berthold Service. The
hardware is defective and, if necessary,
must be checked and replaced.
M109
Lower temperature limit:
Maintenance required
M
The internal temperature of the device
is close to the lower threshold value of
the permissible operating temperature.
M110
Temp LL OOS
S
The internal temperature of the device
is below the lower limit. The correct
function of the device can not be guar-
anteed. It is recommend to have the
device checked by Berthold Service,
even if it seems to work normally.
M111
Temp UL maintenace
M
The internal temperature of the device
is close to the upper limit.
M112
Temp UL OOS
S
The internal temperature of the device
is above the upper limit. The correct
function of the device can not be guar-
anteed. It is recommend to have the
device checked by Berthold Service,
even if it seems to work normally.
M113
Power On Reboot
C
The device was restarted due to un-
known reasons.
M114
Software Reboot
C
The device was restarted by user input.
M115
Extern RTC malfunction
M
Failure of the external real time clock.
Contact Berthold service, if this event
occurs repeatedly.
M116
Corrupt Date
M
The date could not be verified at
startup. Check date and time and set if
necessary.
M107
Intern RTC malfunction
M
Failure of the real time clock. Verify
date and time. Contact Berthold service,
if this event occurs repeatedly.
M108
CPU temp sensor
M
Failure of the device temperature sen-
sor. Contact Berthold service. The
Hardware is defective and needs to be
checked / replaced.
M109
Temp LL maintenance
M
The internal temperature of the device
is close to the lower limit.
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...