10 Maintenance and Repair
LB 470 Level
170
56925BA2 Rev. 03, 11/2017
10.1
Replacing of Fuses
j
DANGER
Danger to life from electric shock!
Replacing of fuses may only be carried out by a qualified electrician.
Please adhere to the relevant safety regulations.
Installation/maintenance may only be carried out if the device has been
de-energised.
Only open the device when free of voltage.
In case of an electric shock, carry out first aid measures and immediately call an
emergency service.
NOTICE
Damage to the device! Short circuit!
The EVU can be damaged if incorrect fuses are used.
Only use fuses which correspond to the fuses on the circuit board of the
module.
Fuses:
•
Master EVU 250V 1A T (5x20 mm)
•
Master EVU 250V TR5 T80mA (Ø 8,5 mm)
•
Slave module 250V 315mA T (5x20 mm)
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...