7 Main Menu Device Setup
LB 470 Level
100
56925BA2 Rev. 03, 11/2017
7.3.3
92B
Calibration
Device settings | Setup | Calibration
The Calibration menu is used for the selection of the calibration and calculation
method and for adaption of the measurement system to the respective environ-
mental conditions, the actual radiation activity and the adaptation of the
background radiation (background level).
NOTICE
Material damage to the device or the system!
Errors in calibration or in the parameter setting can lead to incorrect measure-
ment results. This may possibly lead to loss of production or to damage in the
system.
We encourage you to have the calibration and commissioning performed by
Berthold service.
Fig. 79
Menu "Calibration"
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...