LB 470 Level
9 Troubleshooting
56925BA2 Rev. 03, 11/2017
167
9.2.3
105B
Detector
Code
Message
NAMUR107 Help Text
D501
Detector not found
F
Lost connection to at least one detector.
Check detector settings and connections.
It is recommend to have the device
checked by Berthold Service, even if it
seems to work normally.
D502
Detector comm. error
M
Temporarily lost connection to at least
one detector. Check detector settings
and connections. Contact Berthold ser-
vice, if this event occurs repeatedly.
D503
Detector failure
F
At least one detector register Failure".
Check detector events.
D504
Detector out of spec.
S
At least one detector register "out of
specification". Check detector events.
D505
Detector function check
C
At least one detector register "function
check". Check detector events.
D506
Detector maintenance
M
At least one detector register "mainte-
nance". Check detector events.
D599
Internal program error
F
Internal system failure. Powercycle de-
vice. If the event remains it could be
caused by a defective hardware. Contact
Berthold service.
9.2.4
106B
RS 458 Interface
Code
Message
NAMUR107 Help Text
M699
Internal program error
F
Internal software failure. Restart the de-
vice. Contact Berthold service, if this
event occurs repeatedly.
9.2.5
107B
Process Connection
Code
Message
NAMUR107 Help Text
M701
Current output
F
Deviation of analog output value from
feeback value is too high. Calibrate ana-
log output. Contact Berthold service, if
this event occurs repeatedly.
M702
Current loop open
F
Current output loop open. Check cable
connection.
Содержание Duo Xpert
Страница 2: ......
Страница 4: ......
Страница 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Страница 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Страница 183: ...Unit d valuation D tecteurs LB 47x LB 4700 Informations sur la s curit 56925BA59 Rev No 01 11 2017...
Страница 184: ......
Страница 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Страница 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Страница 208: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017 Evaluation unit Auswerteeinheit...
Страница 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...