Installation and Operation Manual
BEKOKAT
®
V01 /11-046
71 / 88
6.4
Safety temperature monitor (TW) activated
Malfunction image
The safety temperature monitor (TW) of the heating system of the reactor responds,
the temperature at the container tank wall in the upper range is T> + 260 °C.
Cause
Measure
The setpoint of the reactor temperature (in
the d 150 °C) has been changed.
Temperature control on the display with the
Set key, if the setpoint of + 150 °C is set. (see
chapter 5.3.2)
Compressed air flow too low.
Increase flow rate.
Inadmissible high oil concentration, e.g.
due to breakage of the oil separator
cartridge of the compressor.
Replacement of the oil separator cartridge
on the compressor.
BEKOKAT
®
recommissioning procedure.
NOTE
AUTO-RESET
The alarm is reset via AUTO-RESET (see chapter 3.5.7).
The response of the safety temperature monitor (TW) during the heating
phase is not a fault.
This is a normal control status.