Checks/tests
Operating manual AVERO Phoenix
V2.1 E
Page 19 of 35
10.
Checks/tests
In order to ensure a safe use of our AVERO Phoenix lifting bath tub and the protection
of the users and the patients, BEKA Hospitec prescribes an annual safety check.
The execution of the safety checks and maintenance must be documented and proven
on request. Please use your inventory register to this effect.
We recommend a simultaneous maintenance of the device in order to conserve its full
value.
The checks may only be conducted by adequately trained and qualified experts. The
non-observance of this prescription could lead to injuries and jeopardise the safety of
the product.
NOTE
In accordance with the UVV (accident prevention)
regulations of the German employer’s liability insurance
association on stationary equipment which is used in special
locations or installations, an annual check to the DGUV
(German Statutory Accident Insurance Association)
Prescription 3 (BGV A3) must be carried out on the AVERO
Phoenix lifting bath tub.
This check is only prescribed for Germany. In other countries, other requirements
might be applicable.
WARNING
Do not conduct any cleaning, maintenance or test activities
when the bath tub is in use. This could cause danger to the
user and the patient.
measures to be conduct
Before
each
application
/ daily
weekly monthly yearly
Check the level of the disinfectant. (If
available)
X
Clean and disinfect the lifting bath.
X
Check the lifting bath for function.
X
Conduct a visual inspection of all
components, power cables, hoses and
connections.
X
Check the water inlet and clean it if
necessary.
X
We recommend regular maintenance, a
safety check and a check to DGUV
prescription 3.
X
Содержание AVERO Phoenix
Страница 24: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 24 of 35 13 2 Dimensions of the AVERO Phoenix 190...
Страница 25: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 25 of 35 13 3 Dimensions of the AVERO Phoenix 170...
Страница 27: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 27 of 35 13 5 Declaration of Conformity...
Страница 35: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...